Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8062A: CAL ADJ /5100 DATE: 13-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 81 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 8062A instruction manual (P/N 632695, March 1982, Rev. 2, 3/93) # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 51_V8062.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP [32] WARNING 1.003 DISP HIGH VOLTAGE is used or exposed during the performance 1.003 DISP of this calibration. DEATH ON CONTACT may result if 1.003 DISP personnel fail to observe safety precautions. 1.004 DISP Ambient temperature: 21degC - 25degC. 1.004 DISP Relative humidity: less than 80%. 1.004 DISP Allow the UUT 30 min. to stabilize. 1.005 DISP NOTE: The top AC shield should remain installed on the 1.005 DISP instrument while the calibration adjustments are being 1.005 DISP performed. The positions of the trimpots and trimcaps 1.005 DISP are marked on the top AC shield along with a table 1.005 DISP summarizing the calibration procedures. 1.006 DISP The performance of the 8062A AC functions is affected 1.006 DISP by the capacitance between the main PCB and the bottom 1.006 DISP AC shield, which is dependent on the distance between 1.006 DISP them. The distance may vary, depending on whether the 1.006 DISP top cover is installed. To minimize performance 1.006 DISP variations, press down on the AC shield to simulate top 1.006 DISP cover installation while you perform the AC calibration. 1.007 DISP TURN ON UUT. 1.008 HEAD DC VOLTAGE CALIBRATION 1.009 PIC 51_V8062 1.010 DISP Turn R6 fully clockwise (CW). 1.010 DISP Turn R5 fully counterclockwise (CCW). 1.011 RNG 2V 1.012 5100 1.9000V S 2W 1.013 DISP Adj R6 (COARSE) for a reading between 1.9010 and 1.9025 1.014 DISP Adj R5 (FINE) for a UUT reading of 1.9000+/-0.0001V 1.015 5100 * S 1.016 RNG 200mV 1.017 5100 190.0mV S 2W 1.018 DISP Adjust R8 for a reading of 190.00+/-0.01mV 1.019 5100 * S 1.020 HEAD AC VOLTAGE CALIBRATION 1.021 DISP Turn R18 fully clockwise (CW). 1.021 DISP Turn R15 fully counterclockwise (CCW). 1.022 RNG 200mV AC 1.023 5100 100.00mV 200H S 2W 1.024 DISP Adj R15 (COARSE) for a reading between 100.40 and 100.95 1.025 DISP Adj R18 (FINE) for a UUT reading of 100.00+/-0.05mV 1.026 5100 * S 1.027 RNG 2V AC 1.028 5100 1.0000V 10kH S 2W 1.029 DISP Adjust C7 for a reading of 1.0000+/-0.0015V 1.030 END